Service manual 
Touch screen –  
Industrial and laboratory scales 

 

 

 

KERN FKT / IKT / PKT 

                   

 

                    

 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

FKT_IKT_PKT-SH-e-1230 

 

Version 3.0 
1/2012 
GB 

Содержание PKT

Страница 1: ...336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service manual Touch screen Industrial and laboratory scales KERN FKT IKT PKT FKT_IKT_PKT SH e 1230 Version 3 0 1 2012 GB ...

Страница 2: ...s Table of Contents 1 Basic Information 3 2 Introduction 3 3 Menu Map 4 4 Troubleshooting 5 5 Adjustment 6 6 Service Menu 7 6 1 How to Enter the Service Menu 7 6 2 Factory Reset 8 6 3 Factory Adjustment 9 6 4 Linearity Adjustment 11 6 5 Geo Value 13 6 6 Serial Number 14 6 7 Clear Passwords 15 6 8 Remark 15 7 Electronics 16 ...

Страница 3: ...ily regulated field Das Gerät darf nur von geschultem oder beruflich ausgebildetem Fachpersonal z B eichrechtlich anerkannter Instandsetzer repariert werden Die Serviceanleitung ist bindend für Reparaturen Das Gerät muss nach erfolgter Reparatur wieder in den Originalzustand zurückversetzt werden Es dürfen nur Originalersatzteile verwendet werden Hinweis zu konformitätsbewerteten Waagen Reparatur ...

Страница 4: ...4 FKT_IKT_PKT SH e 1230 3 Menu Map ...

Страница 5: ... supply interrupted The weight display changes continually Draught air movement Table floor vibrations The weighing plate is in contact with foreign matter Electromagnetic fields static charging choose different location switch off interfering device if possible The weighing result is obviously incorrect The balance display is not set to zero Adjustment is no longer correct Great fluctuations in t...

Страница 6: ...e technical data in the operating instruction Remark The adjusting of verification balances FKT M and IKT M is only possible when not being blocked by the adjusting switch The switch for this step is located at the bottom of the housing like shown below FKT IKT Display device Verification of the balance is invalid without the seal 1 Turn ON the balance 2 Press the MENU CAL key Then press the Adjus...

Страница 7: ...ight in the centre of the weighing plate 6 After a short time balance will return into weighing mode automatically 7 Remove adjusting weight 6 Service Menu 6 1 How to Enter the Service Menu 1 Turn ON the balance 2 Press the MENU CAL key 3 Press the key then press the Service key 4 The display appears ...

Страница 8: ...ter the password 1 0 1 and confirm with ENTER key 6 2 Factory Reset Reset the application settings to factory settings e g language basic settings weighing technology settings etc with Factory reset key and confirm with ENTER key ...

Страница 9: ...FKT_IKT_PKT SH e 1230 9 6 3 Factory Adjustment 1 Press the Factory adjust key 2 With empty weighing pan press the Zero point adjustment key and wait until the display appears the stability mark ...

Страница 10: ...ect Factory adjust again place the maximum weight on the weighing pan and after that press the Ref Weight adjustment key 4 Balance measuring for approx 1 2 minutes 5 After that the display appears the stability mark 6 Remove the weight ...

Страница 11: ...nd adjustment check balance readings in all measuring points In case of differences make a note 2 Enter the Service Menu and select Linearity 3 The display appears the first linearity point e g 4 Press the Deviation key to enter the difference in digit e g minus 5 digit and confirm with ENTER key ...

Страница 12: ...tolerances press the Corr point key and inscribe the deviation like described above 8 To return to normal weighing mode press the ENTER key Information If there is a positive difference e g 2000 6g 6 digit you should subtract this difference to the stored value e g 2000 0g 6d If there is a negative difference e g 1999 4g 6 digit you should add this difference to the stored value e g 2000 0g 6d ...

Страница 13: ...ce of location only if the balance has not already been adjusted to the location in the factory To receive accurate measuring values it is recommended to adjust the balance with external weights according to chapter 5 Adjustment In this category you can set the value of gravitational acceleration without external weights and thus adapt the scale to local gravitational conditions Please only change...

Страница 14: ...egory you can view or change the balance serial number The number was assigned to it in the factory and should never be altered It is only necessary to change this number after replacing the main board The serial number is given on the balance model label ...

Страница 15: ...r all passwords device code and configuration code Select Clear Passwords and confirm with ENTER 6 8 Remark Please notice the categories Temperature adjust Creep and Startup creep were configured and used only in our production line Therefore it is not recommended to use them ...

Страница 16: ... version D 2 Out 2 5 4 2 3 1 Out 3 3 Performance data Vcemax 35 V DC Icmax 80 mA DC 5 Pin 4 is not allocated IN EX IN EX REF SENSE OUT SIG OUT SIG Load cell Touch JTAG LCD Lightening USB REF SENSE Temp sensor 9 10V Ground PTB Lock Ground Batt Power supply Accumulator Common INP OUT 3 OUT 2 OUT 1 RXD TXD GND RXD TXD GND COM 1 COM 2 Optocoupler ...