MBD-BA-e-1710
12
11 Assistance in case of minor failures
In case of any disturbance in the programme, switch off the scales and disconnect
them from power supply for a while. Then restart the weighing process.
Assistance:
Disturbance
Possible cause
The weight indication does
not appear.
•
The scales are off.
•
Improperly inserted or discharged batteries.
•
No batteries.
The weight indication is
constantly changing.
•
Draught/movements of air.
•
Table/floor vibration.
•
The weighing pan is in contact with solids or is in-
correctly installed.
•
Electromagnetic fields / static charges (choose an-
other location — turn off the device causing the
disturbance, if possible).
The weighing result is obvi-
ously incorrect.
•
The weight indication was not cleared.
•
High temperature fluctuations.
•
The warm-up time was not sufficient.
•
Electromagnetic fields / static charges (choose an-
other location — turn off the device that causes the
disturbance, if possible).
In case of any other error messages, you must restart the scales. Contact the manu-