YDB-03-BA-e-2114
27
5.2 Series KERN TACS, TACJ
Install the density determination module – see ch. 4.3.2
5.2.1 Parameter setting
1. Selection of application
Go to the menu:
In the weighing mode, press the
MENU
button twice.
Press the navigation buttons (
) until
“APL.FUNC” appears.
Confirm by pressing
.
Press the navigation buttons (
) until
“SG” appears.
Confirm by pressing
TARE
. The display
shows “SET” and the current setting.
By pressing
TARE
, select between the
settings “SG” and “S.SG” (mode:
“Determination of density of solids”). The
current setting is marked by the
stabilization indicator.
2. Entry of reference liquid parameters
Confirm by pressing
TARE
. The display
shows “SET” and the currently set
reference liquid (e.g. water).
Press the navigation buttons (
) until
the required reference liquid appears.
Содержание KERN YDB-03
Страница 7: ...YDB 03 BA e 2114 7 Adapter 5 items Allen wrench screw ...
Страница 8: ...8 YDB 03 BA e 2114 2 Dimension mm Frame ...
Страница 9: ...YDB 03 BA e 2114 9 Universal immersion basket for floating and descending solid matter Sample dish ...
Страница 10: ...10 YDB 03 BA e 2114 Platform Beaker ...
Страница 90: ...90 YDB 03 BA e 2114 Back to menu weighing mode press ON OFF key repeatedly ...