Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display , then press
.
„PASS“ will be displayed shortly.
After the adjustment the balance will carry out a self-test.
Remove adjusting weight
during
selftest, the appliance will
return into weighing mode automatically. An adjusting error
or incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.
6.10 Verification
UU
General hints:
According to EU directive 90/384/EEC balances must be officially verified if they are
used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purpose.
d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
Verification notes
An EU type approval exists for balances described in their technical data as
verifiable. If a balance is used where obligation to verify exists as described above, it
must be verified and re-verified at regular intervals.
Re-verification of a balance is carried out according to the respective national
regulations. The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Balances with obligation to verify must be taken out of operation if:
•
The weighing result of the balance is outside the error limit.
Therefore, in
regular intervals load balance with known test weight (ca. 1/3 of the max. load)
and compare with displayed value.
•
The reverification deadline has been exceeded.
GAB-P-BA-e-1411
19