10
FCF-BA-e-1611
Start balance by pressing
.
Press and hold down
, the display shows "
CAL
" briefly. After that the
size of the adjustment weight appears flashing in the display.
Now set the adjusting weight in the centre of the weighing plate.
Press
. Short time later there appears „
CAL F
“, then the automatic
return to the weighing mode. In the display there appears the value of the
adjustment weight. Adjustment has now been completed successfully.
An error during adjustment or the use of an incorrect adjusting weight will
result in an error message „
CAL E
“. Repeat adjustment.
Keep the adjustment close to the balance. Daily control of the weighing exactness is
recommended for quality-relevant applications.