- 14 -
EOB-B_EOE-B_EOS-B-SH-e-1911
Service menu
Turn on the scale
and press and hold the
until the countdown sequence, then
release the
and the scale will show
LF 1
Weight calibration
LF 2
Spec Calibration
LF 3
Gravtational (G value) pRe-Calibration
CAL switch hs to be on
Password is required to use when CAL switch is OFF
LF 1 Weight Calibration :
Calibration can be done with any weight, but the weight should not be less than 1/100 of the max.
capacity and not to exceed the max. capacity
Enter the weight calibration will show the last time the weight calibration value, The following are
two sections of the calibration and single calibration of the operation steps
Two sections of the calibration:
Press
to enter
Confirm display of [ECF 1] with
. [CAL Z] is displayed.
Acknowledge with. [--------] followed by [LOAd
ii
] will be displayed
briefly. Then the display to input the recommended adjustment
weight is displayed
The left digit flashes.
Enter the value of the adjustment weight as follows:
Switch to the next digit to the right with
.
Increase number with
.
Confirm the value entered with
.
(example)
Place the required weight mass onto the scale as indicated on
display and press the
↵
key to calibrate the scale
(Press the ESC key to back to menu LF 1)
(example)
Show stop for 2 seconds