466-BA-defsi-0016 13
3.5 Net-total
weighings
Useful when a mixture of several components will be weighed in a tare cup, and
finally the sum weight of all components weighed in will be needed for direction (net-
total, i.e. without the weight of the tare cup).
Example:
Tare cup on weighing plate, tare on “
0
“
by pressing the
TARE
-key.
Weigh component
n
, and tare back to “
0
” with the
ON/M
key. The activation of the
memory is shown by a triangle on the far left of the display.
Weigh component
o
, by pressing the
ON/M
key the net total is displayed, this means
the sum of component
n
and
o
. Tare to “
0
” by pressing the
ON/M
key.
Weigh component
p
, by pressing the
ON/M
key the net-total is displayed, this means
the sum of the components
n
and
o
and
p
.
If necessary fill the formula container up to the desirable level. To return to the
weighing mode press the
TARE
key.
3.6 Adjusting
(CAL)
The accuracy of the balance can be controlled and set at any time by means of the
enclosed adjusting weight.
Consider stable environmental conditions, see ”Important Information”. A short
warming up time of approx. 1 minute for stabilisation will be useful.
Press
CAL/MODE
key approx. 3 seconds, the exact weight of the adjusting weight
will appear flashing in the display.
Press
OFF
key, “
AL 0
” will appear.
Place the adjusting weight on the centre of the weighing plate.
“
AL F
” will appear somewhat later, and then the automatic return to the normal
weighing mode will follow. The weight of the adjusting weight will appear in the
display.
It is possible to break off the adjusting programme with
OFF
key.
In case of an adjusting error or a wrong adjusting weight “
AL E
” appears in the
display. Press
OFF
key and repeat the adjusting procedure.
Keep the adjusting weight near the balance for daily direction of the balance’s
accuracy.