BOP HIPWR 091719
3-61
3.5.3.4.6 CONFIGURE INDEX
To configure the index number manually, press
%
from the power-up screen, then use
Y
and
U
keys until the display shows the current LAN settings (LAN SETTINGS highlighted). To
modify, press
!
to enter the LAN Interface Setting submenu (see Table 3-18). Under Host,
highlight Index value and press
!
to modify the setting. Enter up to 14 ASCII Characters and
press
$
to apply, then press
$
to save for power-up.
3.5.3.5
BIT 4882 COMPATIBILITY.
The BOP 1KW is not directly compatible with the older linear BOP models using a BIT 4882 for
digital remote control, as there are differences in behavior between the two: In response to the
command line:
outp off;:volt 10
the BIT 4882 would set the output to 10 volts, while the
BOP 1KW will remain in the output OFF state until
OUTP ON
is sent. The closest approximation
to using the BOP 1KW to mimic the BOP/BIT 4882 is to use
OUTP:CONT ON
and
SYST:SET
CM0
commands to configure the unit. The BOP 1KW response to the command line
outp
off;:volt 10
would then be to set the output to OFF for 100 mS, then return to ON.
3.5.3.6
BIT 4886 COMPATIBILITY
Although the SYST:SET command can be used to configure the BOP 1KW the same way the
BIT 4886 is configured, the BOP 1KW is not directly compatible with the older linear BOP mod-
els using a BIT 4886 for digital remote control. There are differences in behavior between the
two in handling LIST commands, status reporting and trigger operation as described below.
3.5.3.6.1 LIST COMMAND DIFFERENCES IN BIT 4886
The LIST commands of the older BOP/BIT 4886 are the same as the BOP 1K except as follows:
•
LIST:DWELL - The BOP/BIT 4886 range is from 0.0005 to 5.0 Seconds while the BOP
1KW range is from 0.000091 to 0.034 Seconds. The BOP 1KW is faster but has less
range than the BOP/BIT 4886.
•
LIST:CLEAR sets the LIST:COUNT to 0 in the BOP 1KW but LIST:COUNT is set to 1 in
the BOP/BIT 4886.
•
The BOP 1KW has additional LIST commands not found in the BOP/BIT 4886.
3.5.3.6.2 STATUS REPORTING DIFFERENCES IN BIT 4886
The status register responses from the BOP 1KW and the BOP BIT/4886 are very different. If
these commands are used, then changes will be required in the user’s program. Commands
affected are:
SYST:OPER
,
SYST:QUES
,
*ESR
and
*STB
.
3.5.3.6.3 TRIGGER OPERATION DIFFERENCES IN BIT 4886
The BOP/BIT 4886 only supports
*TRG
while the BOP 1KW supports
TRIGGER:SOURCE
EXTERNAL
,
BUS
and
IMMEDIATE
. The command
TRIG:SOUR BUS
must be sent before using
trigger in the BOP 1KW.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......