DON’T GAMBLE WITH YOUR SAMPLE
CIP RECIRCULATOR
USER MANUAL
Страница 1: ...DON T GAMBLE WITH YOUR SAMPLE CIP RECIRCULATOR USER MANUAL ...
Страница 2: ... User Manual and states the major changes between ver sions This version list is introduced in September 2017 Version Version date Major changes from previous versions 1 September 2017 Complete revision and new layout 2 April 2019 Replaces preliminary version 1 with more text improved instruc tions and more pictures ...
Страница 3: ... Svendborg Denmark TYPE CIP Recirculator YEAR OF INTRODUCTION 2017 YEAR OF REVISED DESIGN 2019 LAST UPDATED 2019 The English version of this Manual is the governing version and it is the only authorized version Consequently KEOFITT cannot be held liable for other versions including translations of this Manual ...
Страница 4: ... 2 CIP RECIRCULATOR DESCRIPTION 6 3 INITIAL CLEANING 7 4 CHOOSING AND PREPARING A CLEANING SOLUTION 8 5 PRE SAMPLING CLEANING 9 5 1 Connecting the Recirculator 9 5 2 Flushing 9 5 3 CIP Recirculation 10 6 SAMPLING 11 7 POST SAMPLING CLEANING 12 8 ADDITIONAL BARRIER 13 9 RE USE OF CLEANING SOLUTION 14 10 ALTERNATIVE CLEANING AGENTS 15 ...
Страница 5: ... is based on the operator circulating a cleaning agent from the bot tle through the valve chamber and back to the bottle by means of a hand operated bulb pump 1 1 ALCOHOL AS CLEANING AGENT By default this manual is based on alcohol 70 as cleaning agent since it is widely used in breweries For this application alcohol has the advantage that it will act both as a cleaning agent and as a disinfect an...
Страница 6: ... valve 5 Safety check valve Secures a uni directional flow from the bottle through the supply hose to the valve to be cleaned Requires a given over pressure to open at all 6 Inlet connector Fits hose piece connectors on Keofitt valves To be connected to the sampling valve s inlet port upper port 7 Outlet connector Fits hose piece connectors on Keofitt valves To be connected to the sampling valve s...
Страница 7: ...ap and the associated hoses and connectors with alcohol or another appropriate detergent and disinfectant NB Perform these operations in a clean controlled environment The required hygienic level must be determined individually depending the actual application Once washed disinfected keep the Recirculator in a suitable protective packaging until next use ...
Страница 8: ...ution and refit the screw cap Make sure the long suction tube is inserted well below liquid level 4 Provide a small container to place the caps into and fill it with the alcohol solu tion Alcohol is a good detergent for water soluble and alcohol soluble substances and at the same time the alcohol acts as a disinfectant Always use a 70 alcohol solution consisting of 70 ethanol or isopropanol and 30...
Страница 9: ... Connect Recirculator s supply hose to the valve s top port The mentioned caps plugs could be either the standard rubber caps usually suspended on a chain from the valve or separate rubber caps or steel plugs see www keofitt dk for further information 5 2 FLUSHING 8 Squeeze the bulb pump completely 3 4 times to rinse the valve chamber The alcohol is flushed to drain or any other waste collection m...
Страница 10: ...through the valve chamber 5 10 times When the Recirculator s two hoses are connected to the valve the CIP recirculation may start NB It is important to squeeze the rubber bulb pump rather hard in order to obtain an adequate liquid velocity which will mechanically improve the cleaning effect The adequate number of pumping actions is individual from application to application and must be determined ...
Страница 11: ...apped quickly once flushed in order to prevent airborne contaminants from entering the valve chamber Instead of using rubber caps suspended on a chain from the valve you may use disinfected re usable caps rubber or steel which are brought to the sampling point in a container filled with alcohol This is a common practice in some industries It is also in some industries common practice to flush with...
Страница 12: ...ount of product being circulated during the CIP process steps 9 10 At the same time you should clean and disinfect the caps and make them ready to use for plugging the ports unless you use new re useable caps in which case you discard the used caps 19 Connect return hose to lower port 20 Squeeze hand pump 2 3 times to circulate the cleaning agent in the sampling valve This is a step similar to Ste...
Страница 13: ...primary barrier being the valve s mem brane Finally the Recirculator must be disconnected 23 Disconnect the supply hose and lift it a little from the top port while at the same time squeeze the bulb pump slightly to allow alcohol to flow into the valve chamber 24 For obtaining an additional barrier make sure a surface of alcohol is visible inside the top port If not visible you may add some alcoho...
Страница 14: ... SOLUTION If pre CIP flushing removes all visible residues the amount of product that goes into the bottle is very small which may justify re use of the alcohol for the subsequent cleaning processes This must be deter mined by the user depending on the actual application ...
Страница 15: ...ry much like what is described above for alcohol except for the fact that you may need 2 or 3 bottles One bottle for the alkali One bottle for the acid only if acid cleaning is also necessary One bottle for rinse water letting rinse water flow to drain NB This User Manual does not recommend one or the other cleaning agent or cleaning proce dure as they may differ a lot from one industry to another...
Страница 16: ...vendborg Denmark Phone 45 6316 7080 Fax 45 6316 7081 info keofitt dk www keofitt dk Keofitt reserves the right to change technical data without notice For complete set of updated data sheets and manuals for Keofitt products please refer to our web page www keofitt dk ...