Models 2510 and 2510-AT User’s Manual
Getting Started
1-25
Table 1-8
Resistance configuration menu
Configuration menu item
Description
FUNCTION
Select DC resistance function.
AC-OHMS
Select AC resistance function.
TEMP-CONTROL
PROTECTION
CONTROL
LOLIM
HILIM
Program DC resistance parameters.
Configure resistance protection.
ENABLE or DISABLE resistance protection.
Program lower resistance limit.
Program higher resistance limit.
SENSOR-TYPE
THERMISTOR
RANGE
I-SRC
SENSE-MODE
RTD
I-SRC
RANGE
SENSE-MODE
Select sensor type and settings.
Program thermistor settings.
Set thermistor range (100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
).
Set thermistor current (AUTO, 3.3µA, 10µA, 33.3µA, 100µA, 833.3µA,
2.5mA).
Select 2-WIRE or 4-WIRE sense mode.
Program RTD settings.
Set RTD current (AUTO, 3.3µA, 10µA, 33.3µA, 100µA, 833.3µA, 2.5mA).
Set RTD range (100
Ω
or 1000
Ω
).
Select 2-WIRE or 4-WIRE sense mode.
PID
PROPORTIONAL
INTEGRAL
DERIVATIVE
Program PID loop constants for resistance.
Set proportional constant.
Set integral constant.
Set derivative constant.