2-8
Calibration
Models 2510 and 2510-AT
Aborting calibration steps
To abort a calibration step from the front panel, press the EXIT key. To abort a calibration
step via remote, send the :ABORt command.
Front panel calibration
The front panel calibration procedure described below calibrates all functions. Note that
each function is separately calibrated, and the procedure must be performed in the order shown.
Step 1: Prepare the Model 2510 for calibration
1.
Turn on the Model 2510 and the digital multimeter, and allow them to warm up for at
least one hour before performing calibration.
2.
Configure the INPUT terminals for 4-wire sensing as follows:
•
Press CONFIG then T.
•
Select SENSOR-TYPE, then press ENTER.
•
Choose THERMISTOR, then press ENTER.
•
Select SENSE-MODE, then press ENTER.
•
Choose 4-WIRE, then press ENTER.
•
Press EXIT as required to back out of the menu structure and return to normal
display.
3.
Press the MENU key, choose CAL, and press ENTER. Select UNLOCK, and then
press ENTER. The instrument will display the following:
PASSWORD:
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
4.
Use the
¹
and
ƒ
keys to select the letter or number, and use the
ß
and
©
arrow keys to
choose the position. (Press
ƒ
for letters;
¹
for numbers.) Enter the present password on
the display. (Front panel default: 002510.)
5.
Press ENTER to complete the process.
6.
Select EXECUTE, then press ENTER to enter the CAL EXECUTE menu. Table 2-4
summarizes the various menu selections and calibration steps, which are covered in
more detail throughout the procedure.
2510-902-01 (E - Mar 2010)(Service).book Page 8 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...