2-6
Calibration
Models 2510 and 2510-AT
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE
'<password>'
For example, the following command uses the default password:
:CAL:PROT:CODE
'KI002510'
Changing the password
The default password (002510) may be changed from the front panel or via remote as
discussed.
Changing the password from the front panel
Follow the steps below to change the password from the front panel:
1.
Press the MENU key, choose CAL, and press ENTER. The instrument will display the
following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
2.
Select UNLOCK, then enter the password. (Default: 002510.)
3.
Select CHANGE-PASSWORD, and then press ENTER. The instrument will display
the following:
New Pwd: 002510
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
4.
Using the EDIT keys, enter the new password on the display.
5.
Once the desired password is displayed, press the ENTER key to store the new
password.
Changing the password by remote
To change the calibration password by remote, first send the present password, and then
send the new password. For example, the following command sequence changes the password
from the 'KI002510' remote default to 'KI_CAL':
:CAL:PROT:CODE
'KI002510'
:CAL:PROT:CODE
'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTE
If you change the first two characters of the password to something other than “KI”,
you will not be able to unlock calibration from the front panel.
2510-902-01 (E - Mar 2010)(Service).book Page 6 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...