Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-236
2600BS-901-01 Rev. B / May 2013
Also see
Triggering
(on page 3-32)
smuX.trigger.count
This attribute sets the trigger count in the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
1
Usage
triggerCount
= smu
X
.trigger.count
smu
X
.trigger.count =
triggerCount
triggerCount
The trigger count is the number of times the source-measure unit (SMU) will iterate
in the trigger layer for any given sweep
X
SMU channel (for example,
smua.trigger.count
applies to SMU channel A)
Details
During a sweep, the SMU iterates through the trigger layer of the trigger model this many times. After performing
this many iterations, the SMU returns to the arm layer.
If this count is set to zero (
0
), the SMU stays in the trigger model indefinitely until aborted.
Also see
Triggering
(on page 3-32)
smuX.trigger.endpulse.action
This attribute enables or disables pulse mode sweeps.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
1 (smuX.SOURCE_HOLD)
Usage
pulseAction
= smu
X
.trigger.endpulse.action
smu
X
.trigger.endpulse.action =
pulseAction
pulseAction
The pulse mode setting; set to one of the following values (see Details for
definition):
0
or
smu
X
.SOURCE_IDLE
1
or
smu
X
.SOURCE_HOLD
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.action
applies to SMU channel A)
Details
When set to
smu
X
.SOURCE_HOLD
, this attribute disables pulse mode sweeps, holding the source level for the
remainder of the step.
When set to
smu
X
.SOURCE_IDLE
, this attribute enables pulse mode sweeps, setting the source level to the
programmed (idle) level at the end of the pulse.