Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-206
2600BS-901-01 Rev. B / May 2013
smuX.measure.highcrangedelayfactor
This attribute contains a delay multiplier that is only used during range changes when the high-capacitance mode
is active.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
10
Usage
delayFactor
= smu
X
.measure.highcrangedelayfactor
smu
X
.measure.highcrangedelayfactor =
delayFactor
delayFactor
The delay factor; set to a value between 1 and 99
X
SMU channel (for example,
smua.measure.highcrangedelayfactor
applies
to SMU Channel A)
Details
This delay multiplier is only active when the high-capacitance mode is active.
Example
smua.measure.highcrangedelayfactor = 5
Increases the delay used during range changes for
SMU channel A by a factor of 5.
Also see
High-capacitance mode
(on page 3-65)
reset()
(on page 7-160)
setup.recall()
(on page 7-180)
smuX.reset()
(on page 7-216)
smuX.source.highc
(on page 7-221)
smuX.measure.interval
This attribute sets the interval between multiple measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
0 (0 s)
Usage
interval
= smu
X
.measure.interval
smu
X
.measure.interval =
interval
interval
The interval value (in seconds); set to a value between 0 and 1
X
Source-measure unit (SMU) channel (for example,
smua.measure.interval
applies to SMU channel A)