Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-132
2600BS-901-01 Rev. B / May 2013
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to incoming LAN
trigger packets.
Example
digio.trigger[14].stimulus =
lan.trigger[1].EVENT_ID
Route occurrences of triggers
on LAN trigger 1 to digital I/O
trigger 14.
Also see
None
lan.trigger[N].ipaddress
This attribute specifies the address (in dotted-decimal format) of UDP or TCP listeners.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
LAN trigger N reset
Recall setup
Not saved
"0.0.0.0"
Usage
ipAddress
= lan.trigger[
N
].ipaddress
lan.trigger[
N
].ipaddress =
ipAddress
ipAddress
The LAN address for this attribute as a string in dotted decimal notation
N
A number specifying the LAN event number (1 to 8)
Details
Sets the IP address for outgoing trigger events.
Set to
"0.0.0.0"
for multicast.
After changing this setting, the
lan.trigger[
N
].connect()
command must be called before outgoing
messages can be sent.
Example
lan.trigger[3].protocol = lan.TCP
lan.trigger[3].ipaddress = "192.168.1.100"
lan.trigger[3].connect()
Set the protocol for LAN trigger 3 to be
lan.TCP
when sending LAN triggers.
Use IP address
"192.168.1.100"
to
connect the LAN trigger.
Also see
lan.trigger[N].connect()
(on page 7-130)