![Keithley SourceMeter 2600B Series Скачать руководство пользователя страница 857](http://html2.mh-extra.com/html/keithley/sourcemeter-2600b-series/sourcemeter-2600b-series_reference-manual_4085220857.webp)
Appendix E: Status model
Series 2600B System SourceMeter® Instrument Reference Manual
E-16
2600BS-901-01 Rev. C / August 2016
Status Byte Register
The summary messages from the status registers and queues are used to set or clear the appropriate
bits (B0, B1, B2, B3, B4, B5, and B7) of the Status Byte Register. These summary bits do not latch,
and their states (0 or 1) are dependent upon the summary messages (0 or 1). For example, if the
Standard Event Register is read, its register will clear. As a result, its summary message will reset to
0, which will then reset the ESB bit in the Status Byte Register.
The Status Byte Register also receives summary bits from itself, which sets the Master Summary
Status, or MSS, bit.
Figure 172: Status byte and service request (SRQ)