![Keithley SourceMeter 2600B Series Скачать руководство пользователя страница 791](http://html2.mh-extra.com/html/keithley/sourcemeter-2600b-series/sourcemeter-2600b-series_reference-manual_4085220791.webp)
Series 2600B System SourceMeter® Instrument Reference Manual
Appendix B: Calibration
2600BS-901-01 Rev. C / August 2016
B-15
Voltage source accuracy
Follow the steps below to verify that the Series 2600B output voltage accuracy is within specified
limits. To perform this test, you will set the output voltage to each full-range value and measure the
voltages with a precision digital multimeter.
1. With the power off, connect the digital multimeter (DMM) to the Series 2600B output terminals
using 4-wire connections, as shown below.
Figure 145: Connections for voltage verification