Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-321
As an example, to set bit B1 of the operation status TSP-Link overrun enable register, set
status.operation.instrument.tsplink.trigger_overrun.enable =
status.operation.instrument.tsplink.trigger_overrun.LINE1
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B1 and B3, set
operationRegister
to 10 (which is the
sum of 2 + 8).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example 1
status.operation.instrument.tsplink.trigger_overrun.enable =
status.operation.instrument.tsplink.trigger_overrun.LINE1
Uses a constant
to set the line 1
bit of the
operation status
TSP-Link
overrun enable
register.
Also see
(on page E-9)
status.operation.instrument.trigger_timer.*
status.operation.measuring.*
This attribute contains the operation status measuring summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B:
2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B:
6 (All bits set)