![Keithley SourceMeter 2600B Series Скачать руководство пользователя страница 627](http://html2.mh-extra.com/html/keithley/sourcemeter-2600b-series/sourcemeter-2600b-series_reference-manual_4085220627.webp)
Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-288
2600BS-901-01 Rev. C / August 2016
Details
These attributes read or write to the measurement event voltage limit summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is bit B0, and the most significant bit is bit
B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-19). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.measurement.voltage_limit.SMUA
Set bit indicates the enabled VLMT bit for the
SMU A measurement register is set.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.measurement.voltage_limit.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates the enabled VLMT bit for the SMU
B measurement register is set.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event voltage limit summary enable register, set
status.measurement.voltage_limit.enable =
status.measurement.voltage_limit.SMUA
.
In addition to the above constants,
measurementRegister
can be set to the decimal equivalent of
the bit to set. To set more than one bit of the register, set
measurementRegister
to the sum of
their decimal weights. For example, to set bits B1 and B2, set
measurementRegister
to 6 (which is
the sum of 2 + 4).
Example
status.measurement.voltage_limit.enable =
status.measurement.voltage_limit.SMUA
Sets the SMUA bit of the
measurement event voltage limit
summary enable register using a
constant.
Also see
(on page E-8)
status.node_enable
This attribute stores the system node enable register. This attribute is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Status reset
Not saved
0