Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-219
smuX.measure.delayfactor
This attribute stores a multiplier to the delays that are used when
smu
X
.measure.delay
is set to
smu
X
.DELAY_AUTO
.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Not saved
1
Usage
delayFactor
= smu
X
.measure.delayfactor
smu
X
.measure.delayfactor =
delayFactor
delayFactor
The delay factor multiplier
X
Source-measure unit (SMU) channel (for example,
smua.measure.delayfactor
applies to SMU channel A)
Details
The delay factor is only applied when
smu
X
.measure.delay = smu
X
.DELAY_AUTO
.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic
delay.
Setting this attribute to zero disables delays when
smu
X
.measure.delay = smu
X
.DELAY_AUTO
.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay for SMU channel A.
Also see
(on page 2-82)
(on page 7-173)
(on page 7-218)
(on page 7-232)
smuX.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
1
Usage
filterCount
= smu
X
.measure.filter.count
smu
X
.measure.filter.count =
filterCount
filterCount
The number of readings required for each filtered measurement (1 to 100)
X
Source-measure unit (SMU) channel (for example,
smua.measure.filter.count
applies to SMU channel A)