Series 2600B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2600BS-901-01 Rev. C / August 2016
2-31
Setting the sink mode using the front panel
To enable or disable the sink mode from the front panel:
1. Press the
CONFIG
key and then the
SRC
key.
2. Select
V-SOURCE
.
3. Select
SINK-MODE
.
4. Select
ENABLE
or
DISABLE
.
5. Push the
ENTER
key. Sink mode is enabled or disabled, as applicable.
6. Press the
EXIT (LOCAL)
key twice to return to the main display.
Setting the sink mode from the remote interface
To enable or disable sink mode from the remote interface:
The programming example below illustrates how to enable sink mode (to disable, set the attribute to
smua.DISABLE
):
smua.source.sink = smua.ENABLE
Fundamental circuit configurations
The fundamental source-measure configurations for the Series 2600B are shown in the following
figure. When sourcing voltage, you can measure current or voltage (see A: Source V). When sourcing
current, you can measure voltage or current (see B: Source I). See
page 4-20) for detailed information.
Figure 10: Fundamental source-measure configurations
Operation considerations for the ADC
The following paragraphs discuss autozero and NPLC caching.