Series 2600B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2600BS-901-01 Rev. C / August 2016
2-21
The following table contains descriptions of the configuration menus, as well as cross-references to
related information. To select a menu for single SMU instruments, press the
CONFIG
key and then
the front-panel key associated with the menu (see the description column in the following table). For
two-SMU instruments, press the
CONFIG
key, select the appropriate channel (CHANNEL-A,
CHANNEL-B) and then the front-panel key associated with the menu. To select TRIG and STORE
menus on two-SMU instruments, press the
CONFIG
key and then select COMMON instead of
selecting a channel.
To access, press the
CONFIG
key and then:
Options
For more information, see:
SRC
V-source sense, low range;
I-source low range; and
highC mode
MEAS
V and I-measure range, V-
measure sense, low range;
autozero
LIMIT
V-source and I-source
compliance limits
SPEED
Measurement speed (NPLC)
REL
Set relative values
FILTER
Control digital filter
OUTPUT ON/OFF
Set off-state, control digital
I/O
TRIG
Set trigger in, count, interval,
and delay
(on page 3-32)
STORE
Set buffer count and
destination
(on page 4-1)
Setting values
Setting a value
You can adjust a value using either the
Navigation wheel method
or
Numeric entry method
(using
the keypad).
Navigation wheel method:
1. Use the
CURSOR
arrow keys (or turn the navigation wheel ) to move the cursor to the digit that
needs to be changed.
2. Press the navigation wheel or the
ENTER
key to enter edit mode. The EDIT indicator is
illuminated.
3. Rotate the navigation wheel to set the appropriate value.
4. Press the
ENTER
key to select the value or press the
EXIT (LOCAL)
key to cancel the change.
5. To return to the main menu, press the
EXIT (LOCAL)
key.
Numeric entry method:
The numeric entry method may only be used if the numeric keypad is enabled.