Table 8-1
Maximum compliance limits
Model 2601/2602
Model 2611/2612
Source
range
Maximum
compliance
value
Source
range
Maximum
compliance
value
100mV
1V
6V
40V
3A
3A
3A
1A
200mV
2V
20V
200V
1.5A
1.5A
1.5A
100mA
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
3A
40V
40V
40V
40V
40V
40V
40V
40V
6V
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
200V
200V
200V
200V
200V
200V
200V
20V
20V
Series 2600 System SourceMeters Reference Manual
Source-Measure Concepts
8-3
Return to
2600S-901-01 Rev. A / May 2006
Compliance principles
Compliance acts as a clamp. If the output reaches the compliance value, the
SourceMeter will attempt to prevent the output from exceeding that value. This
action implies that the source will switch from a V-source to an I-source (or from
an I-source to a V-source) when in compliance.
As an example, assume the following:
SourceMeter: V
SRC
= 10V; I
CMPL
= 10mA
DUT resistance: 10
Ω
With a source voltage of 10V and a DUT resistance of 10
Ω
, the current through
the DUT should be: 10V/10
Ω
= 1A. However, because the compliance is set to
10mA, the current will not exceed that value, and the voltage across the
resistance is limited to 100mV. In effect, the 10V voltage source is transformed
into a 10mA current source with a 100mV compliance value.