2-4
Calibration
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use
alternate equipment as long as that equipment has specifications comparable to those
listed in the table. For optimum calibration accuracy, test equipment specifications should
be at least four times better than corresponding Model 2425 specifications.
Table 2-1
Recommended calibration equipment
Description
Manufacturer/Model
Specifications
Digital Multimeter
Hewlett Packard
HP3458A
DC Voltage
1
DC Current
1
1V:
10V:
100V:
10
μ
A:
100
μ
A:
1mA:
10mA:
100mA:
1A:
±5.6ppm
±4.3ppm
±6.3ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
Precision Resistor
2
Isotec RUG-Z-1R00-0.1
1
Ω
, ±0.1%, 100W
1
90-day, full-range accuracy specifications of ranges required for various measurement points.
2
Necessary for calibration of 3A current range. Resistor must be characterized to ±300ppm or better using
recommended DMM before calibrating 3A range.
1
Ω
resistor characterization
The 1
Ω
resistor must be characterized to ±300ppm or better before calibrating the 3A cur-
rent range. Use the 4-wire ohms function of the DMM recommended in Table 1-1 to mea-
sure the resistance value, and then use that measured value to calculate the current during
the 3A current range calibration procedure.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......