2400-902-01 (G - Feb 2006)
Model 2400 Service Manual
1-2
Performance Verification
Introduction
Use the procedures in this section to verify that Model 2400 accuracy is within the limits stated
in the instrument’s one-year accuracy specifications. You can perform these verification
procedures:
•
When you first receive the instrument to make sure that it was not damaged during
shipment.
•
To verify that the unit meets factory specifications.
•
To determine if calibration is required.
•
Following calibration to make sure it was performed properly.
WARNING
The information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, which
could cause personal injury or death if contacted. Use standard safety pre-
cautions when working with hazardous voltages.
NOTE
If the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the correct
course of action.
Verification test requirements
Be sure that you perform the verification tests:
•
Under the proper environmental conditions.
•
After the specified warm-up period.
•
Using the correct line voltage.
•
Using the proper test equipment.
•
Using the specified output signal and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•
An ambient temperature of 18-28•C (65-82•F).
•
A relative humidity of less than 70% unless otherwise noted.
Warm-up period
Allow the Model 2400 to warm up for at least one hour before conducting the verification
procedures. If the instrument has been subjected to temperature extremes (those outside the
ranges stated above), allow additional time for the instrument’s internal temperature to
stabilize. Typically, allow one extra hour to stabilize a unit that is 10•C (18•F) outside the
specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Содержание SourceMeter 2400
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Страница 29: ...2 Calibration...
Страница 51: ...3 RoutineMaintenance...
Страница 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Страница 55: ...4 Troubleshooting...
Страница 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Страница 69: ...5 Disassembly...
Страница 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Страница 77: ...6 ReplaceableParts...
Страница 91: ...A Specifications...
Страница 111: ...B CommandReference...
Страница 123: ...C CalibrationPrograms...
Страница 130: ...C 8 Calibration Programs...
Страница 134: ......
Страница 136: ......