Series 2200 Programmable DC Power Supplies Reference Manual
Section 2: General operation
2200S-901-01 Rev. C / March 2016
2-33
Programming commands
To reset the power supply:
*RST
To set the trigger source to be internal:
TRIGGER:SOURCE:BUS
To configure the list:
SOURCE:LIST:STEP 5
SOURCE:LIST:MODE CONT
SOURCE:LIST:COUT 2
SOURCE:LIST:VOLT 1, 2V
SOURCE:LIST:VOLT 2, 5V
SOURCE:LIST:VOLT 3, 1V
SOURCE:LIST:VOLT 4, 3V
SOURCE:LIST:VOLT 5, 7V
SOURCE:LIST:CURR 1, 5A
SOURCE:LIST:CURR 2, 5A
SOURCE:LIST:CURR 3, 1A
SOURCE:LIST:CURR 4, 3A
SOURCE:LIST:CURR 5, 1A
SOURCE:LIST:WIDTH 1, 0.001S
SOURCE:LIST:WIDTH 2, 0.001S
SOURCE:LIST:WIDTH 3, 1S
SOURCE:LIST:WIDTH 4, 0.001S
SOURCE:LIST:WIDTH 5, 0.001S
SOURCE:LIST:SAV 3
To run the list:
OUTPUT ON
SOURCE:FUNC:MODE LIST
*TRG
Default setup
The following table lists the settings that are restored when you return the power supply to default
settings.
Default setup
Menu or system
Defaults
VOLT:PROT
MAX
VOLT:PROT:STAT
OFF
OUTP
OFF
VOLT
1 V
VOLT:RANG
MAX
CURR
0.1 A
OUTP:TIM
60
OUTP:TIM:STAT
OFF