Introduction

This guide contains information on S46T relay installation and system configuration that is necessary after 
installing new relays. To install a relay, refer to 

“Relay kits” on page 1

 for a description of the kits and then 

follow the applicable installation instructions.

WARNING:

The information in this section is only intended for qualified service 
personnel. Do not attempt to replace parts or otherwise service the 
equipment unless you are qualified to do so.

Relay kits

Table 1

 summarizes available relay kits and part numbers.

WARNING:

Use only the parts specified in 

Table 1

 of this guide when servicing the 

equipment. Use of improper parts may expose the operator to 
hazardous voltages that could result in personal injury or death.

Table 1
S46T relay kits

Relay kit 

model number

Description

Relay

location

Kit parts summary

(Quantity)

Keithley 

part number

Installation 

reference

S46T-SPDT-KIT

18GHz SPDT

1 to 8  SPDT un-terminated relay (1) 

S40-0154 

“SPDT relay 1–

8 installation” 

on page 3

SPDT spacer (1) 

S40-SE05-307 

Connecting cable (1) 

SYS46-311C 

S46T-SP4T-KIT

18GHz SP4T A to D  4-40 ×1/4 PPHSEM screws (2)  4-40X1/4PPHSEM 

“Relay A–D 

installation” on 

page 5

4-40 ×3/8 PFH screws (4)

4-40x3/8PFH

SP4T relay (1)

RL-282-2

Ribbon cable (1) 

CA-286-1 

Mounting plate (1)

S40-N410-312

S46T-SP6T-KIT

18GHz SP6T  A to D  4-40 ×1/4 PPHSEM screws (2)  4-40X1/4PPHSEM 

“Relay A–D 

installation” on 

page 5

4-40 ×3/8 PFH screws (4)

4-40x3/8PFH

SP6T un-terminated relay (1)

RL-282-1

Ribbon cable (1)

CA-286-1

Mounting plate (1)

S40-N410-312

S46T-SPDT-KIT-26

26.5GHz 

SPDT

1 to 8  SPDT un-terminated relay (1) 

RL-308

“SPDT relay 1–

8 installation” 

on page 3

SPDT spacer (1) 

S40-SE05-307 

Connecting cable (1) 

SYS46-311C

S46T-SP4T-KIT-26

26.5GHz 

SP4T

A to D  4-40 ×3/8 PFH screws (4)

4-40x3/8PFH

“Relay A–D 

installation” on 

page 5

SP4T un-terminated relay (1)

RL-312

Mounting plate (1)

S40-N410-312

Ribbon cable (1) 

S46-U100-350A

Содержание S46T-SPDT-KIT

Страница 1: ...able 1 S46T relay kits Relay kit model number Description Relay location Kit parts summary Quantity Keithley part number Installation reference S46T SPDT KIT 18GHz SPDT 1 to 8 SPDT un terminated relay 1 S40 0154 SPDT relay 1 8 installation on page 3 SPDT spacer 1 S40 SE05 307 Connecting cable 1 SYS46 311C S46T SP4T KIT 18GHz SP4T A to D 4 40 1 4 PPHSEM screws 2 4 40X1 4PPHSEM Relay A D installatio...

Страница 2: ...PHSEM screws 2 4 40X3 16PPHSEM Relay A D installation on page 5 SP4T terminated relay 1 RL 310 Ribbon cable 1 S40 SV11 350 S46T SP6T KIT 26T 26 5GHz SP6T A to D 4 40 3 16 PPHSEM screws 2 4 40X3 16PPHSEM Relay A D installation on page 5 SP6T terminated relay 1 RL 290 1 Ribbon cable 1 S40 SV11 350 S46T XFER KIT 18GHz XFER A to D M3X5PPH screws 4 M3X5PPH Relay A D installation on page 5 XFER un termi...

Страница 3: ...ated SPDT relay installation 5 Install the 2 screws removed in step 2 in the new relay column 6 Thread the red and black wires from the relay through the PVC Tubing PVC tubing TX 52 7 is part of SYS46 311C not shown The TX 52 7 may be cut to a length that will allow the relay assembly to be removed without removing the top cover 7 Referring to Figure 2 solder the two control wires to the terminals...

Страница 4: ...ht most relay column Either relay type Terminated or Un Terminated may be installed in any of the 8 locations S40 SE05 307 Un terminated SPDT relay Relay spacer screws 2 per relay column Un terminated SPDT relay installation Relay 1 2 3 4 5 6 7 8 Connector J18 J19 J20 J21 J22 J23 J24 J25 Relay A B C D Connector J5 J6 J7 J8 RELAY 1 8 Connectors J18 J25 RELAY A D Connectors J5 J8 R B J 6 R C J 7 R D...

Страница 5: ...r screws removed in step 1 4 Plug the ribbon connector into the appropriate circuit board connector see Figure 4 5 Install the top cover and then configure the system as covered in System configuration on page 6 of this guide Figure 5 Connecting control wires to relays A D Plug in cable securing with the supplied screws 26 5GHz Un terminated Relay S40 N410 309 XFER Relay Spacer S40 N410 313 BN BN ...

Страница 6: ... location is not populated For example if RELAY A and RELAY B are SP6T RELAY 1 and RELAY 2 are SPDTs and all other locations are not populated the command syntax would be ROUT CONF CPOL 6 6 0 0 1 1 0 0 0 0 0 0 S46T MSPDT KIT and S46T XFER KIT are programmed with a value of 3 not the number of throws as other relays The MSPDT top relay uses the first channel assignment for relay s A D the second ch...

Отзывы: