Troubleshooting
75
5
5
5
5
5
5
5
5
5
Table 4: Troubleshooting Problems
Symptom
Possible Cause
Possible Solution
Module does not
respond.
The module
configuration is
incorrect.
Check the configuration of your device
driver; see the instructions in the
KUSB-3100 Getting Started Manual
.
The module is
damaged.
Contact Keithley for technical support;
refer to
.
Intermittent
operation.
Loose connections or
vibrations exist.
Check your wiring and tighten any loose
connections or cushion vibration sources;
see the instructions in the
KUSB-3100
Getting Started Manual
.
The module is
overheating.
Check environmental and ambient
temperature; consult the module’s
specifications on
of this manual
and the documentation provided by your
computer manufacturer for more
information.
Electrical noise exists.
Check your wiring and either provide
better shielding or reroute unshielded
wiring; see the instructions in the
KUSB-3100 Getting Started Manual
.
Device failure
error reported.
The module cannot
communicate with the
Microsoft bus driver
or a problem with the
bus driver exists.
Check your cabling and wiring and tighten
any loose connections; see the
instructions in the
KUSB-3100 Getting
Started Manual
.
The module was
removed while an
operation was being
performed.
Ensure that your module is properly
connected; see the instructions in the
KUSB-3100 Getting Started Manual
.
Содержание KUSB-3100
Страница 10: ...Contents x...
Страница 14: ...About this Manual xiv...
Страница 15: ...1 1 Overview Features 2 Supported Software 3...
Страница 18: ...Chapter 1 4...
Страница 46: ...Chapter 2 32...
Страница 47: ...33 3 Supported Device Driver Capabilities...
Страница 86: ...Chapter 4 72...
Страница 87: ...73 5 Troubleshooting General Checklist 74 Service and Support 77...
Страница 100: ...Chapter 6 86...
Страница 101: ...87 A Specifications...
Страница 112: ...Appendix A 98...
Страница 113: ...99 B Connector Pin Assignments...
Страница 115: ...101 C Reading from or Writing to the Digital Registers...