Section 2: General operation
Model 2450 Interactive SourceMeter® Instrument Reference Manual
2-66
2450-901-01 Rev. B/September 2013
You cannot combine the command sets.
As delivered from Keithley Instruments, the Model 2450 is set to work with the Model 2450 SCPI
command set.
If you choose the SCPI2400 command set, you will not have access to some of the extended ranges
and other features that are now available using the SCPI command set. In addition, some
Series 2400 code will work differently in the Model 2450 than it did in the earlier instrument. See
Model 2450 in a Model 2400 application
(on page D-1) for information about the differences.
To set the command set from the front panel:
1. Press the
MENU
key.
2. Under System, select
Settings
.
3. Select the button next to Command Set.
4. Select the command set.
5. You are prompted to reboot.
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
To change to the SCPI 2400 command set from a remote interface:
Send the command:
*LANG SCPI2400
Reboot the instrument after changing the command set.
To verify which command set is selected:
Send the command:
*LANG?
System information
You can get the serial number, firmware version, firmware build, detected line frequency, calibration
verify date, calibration adjust date, and calibration adjust count information from the instrument.
To view the system information from the front panel:
1. Press the
MENU
key.
2. Under System, select
Information
. The system information displays.
3. To return to the Home page, select
Home
.