Section 8: TSP command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
8-222
2450-901-01 Rev. B/September 2013
trigger.timer[N].start.overrun
This attribute indicates if an event was ignored because of the event detector state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Not saved
Not applicable
Usage
state
= trigger.timer[
N
].start.overrun
trigger.timer[
N
].start.overrun =
state
state
The trigger overrun state (
true
or
false
)
N
Trigger timer number (1 to 4)
Details
This attribute indicates if an event was ignored because the event detector was already in the detected state
when the event occurred.
This is an indication of the state of the event detector built into the timer itself. It does not indicate if an overrun
occurred in any other part of the trigger model or in any other construct that is monitoring the delay completion
event. It also is not an indication of a delay overrun.
Example
print(trigger.timer[1].start.overrun)
If an event was ignored, the output
is
true
.
If the event was not ignored, the
output is
false
.
Also see
(on page 8-220)
trigger.timer[N].start.seconds
This attribute configures an alarm or a time in the future when the timer will start.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
0
Usage
time
= trigger.timer[
N
].start.seconds
trigger.timer[
N
].start.seconds =
time
time
The time in seconds (0 to 2147483647)
N
Trigger timer number (1 to 4)
Example
trigger.timer[1].start.seconds = localnode.gettime() + 30
trigger.timer[1].enable = trigger.ON
Set the trigger timer to start 30 seconds from the time when the timer is enabled.
Also see
None