Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-151
:TRIGger:DIGital<n>:IN:OVERrun?
This command returns the event detector overrun status.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:DIGital<n>:IN:OVERrun?
<n>
Digital I/O trigger line (1 to 6)
Details
This command returns the event detector overrun status as
0
(false) or
1
(true).
If this is true, an event was ignored because the event detector was already in the detected state when the event
occurred.
This is an indication of the state of the event detector built into the line itself. It does not indicate if an overrun
occurred in any other part of the trigger model or in any other detector that is monitoring the event.
Example
TRIG:DIG1:IN:OVER?
Returns
0
if no overruns have occurred or
1
if an overrun has occurred.
Also see
(on page 6-19)
(on page 3-62)
:TRIGger:DIGital<n>:OUT:LOGic
This command sets the output logic of the trigger event generator to positive or negative for the specified line.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
POS
Usage
:TRIGger:DIGital<n>:OUT:LOGic <logicType>
:TRIGger:DIGital<n>:OUT:LOGic?
<n>
Digital I/O trigger line (1 to 6)
<logicType>
The output trigger logic of the trigger event generator:
POSitive
or
NEGative