Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-92
2450-901-01 Rev. B/September 2013
:STATus:PRESet
This command resets all bits in the status model.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:STATus:PRESet
Details
This function clears the event registers and the enable registers for operation and questionable. It will not clear
the enable status request enable (
*SRE
) to standard enable (
*ESE
).
Preset does not affect the event queue.
The Status Event Status Register is not affected by this command.
Example
STAT:PRES
Resets the registers.
Also see
(on page C-1)
:STATus:QUEStionable:CONDition?
This command reads the Questionable Condition Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:QUEStionable:CONDition?
Details
This command reads the contents of the Questionable Condition Register, which is one of the Questionable
Event Registers.
For detail on interpreting the value of a register, see
(on page C-15).
Example
:STAT:QUES:COND?
Reads the Questionable Condition Register.
Also see
None