Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-78
2450-901-01 Rev. B/September 2013
:SOURce[1]:LIST:<function>
This command allows you to set up a list of custom values for a sweep.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:SOURce[1]:LIST:CURRent <list>
:SOURce[1]:LIST:CURRent?
:SOURce[1]:LIST:VOLTage <list>
:SOURce[1]:LIST:VOLTage?
<list>
Current:
−
1.05 A to 1.05 A
Voltage:
−
210 V to 210 V
See
Details
Details
This command defines a list of up to 100 source values for a source list. This list is used by the
:SOURce[1]:SWEep:<function>:LIST
command to define the source values for the sweep.
When you start the sweep, the instrument sequentially sources each current or voltage value in the list. A
measurement is performed at each source level.
If there is an existing list, it is replaced by the new list.
When you send this command, the instrument creates a source configuration list named
CurrCustomSweepList
if the function is set to current or
VoltCustomSweepList
if the function is set to
voltage.
To add source values to an existing list, use the
:SOURce[1]:LIST:<function>:APPend
command.
Example
*RST
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SENS:CURR:RSEN OFF
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SOUR:VOLT:ILIM 1
SOUR:LIST:VOLT 1, 5, 1, 5, 1, 5
SOUR:SWE:VOLT:LIST 1, 0.2
INIT
*WAI
TRAC:DATA? 1, 6, "defbuffer1", SOUR, READ
This example will source 1 V, 5 V, 1 V, 5 V,
1 V, 5 V and measure the resulting current
at each voltage point. The time duration of
each voltage point is 200 ms.
Also see
:SOURce[1]:LIST:<function>:APPend
(on page 6-79)
:SOURce[1]:SWEep:<function>:LIST
(on page 6-85)