Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-27
:DISPlay:USER<n>:TEXT[:DATA]
This command defines the text that is displayed on the front-panel User Display screen.
Type
Affected by
Where saved
Default value
Command only
Power cycle
Not applicable
Not applicable
Usage
:DISPlay:USER<n>:TEXT[:DATA] <textMessage>
<n>
The line of the User Display on which to display text:
•
Top line:
1
•
Bottom line:
2
<textMessage>
String that contains the message; up to 20 characters for
USER1
and
32 characters for
USER2
Details
USER1
sets the text for the top line of the User Display.
USER2
sets the text for the bottom line.
These commands define text messages for the User Display swipe screen.
If you enter too many characters, the instrument displays an error message and shortens the message to fit.
Example
DISP:CLE
DISP:SCR USER
DISP:USER1:TEXT "Batch A122"
DISP:USER2:TEXT "Test running"
Clear the User Display screen and switch to
display the User Display screen.
Set the first line to read "Batch A122" and
the second line to display "Test running".
Also see
(on page 6-26)