Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-342
DMM7510-901-01 Rev. B / May 2015
trigger.tsplinkout[N].logic
This attribute defines the trigger output with output logic for a trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
TSP-Link line
N
reset
Configuration script
trigger.LOGIC_NEGATIVE
Usage
logicType
= trigger.tsplinkout[
N
].logic
trigger.tsplinkout[
N
].logic =
logicType
logicType
The output logic of the trigger generator:
•
Assert a TTL-high pulse for output:
trigger.LOGIC_POSITIVE
•
Assert a TTL-low pulse for output:
trigger.LOGIC_NEGATIVE
N
The trigger line (1 to 3)
Details
This attribute controls the logic that the output trigger generator uses on the given trigger line.
The output state of the digital I/O line is controlled by the trigger logic, and the user-specified output
state of the line is ignored.
Example
tsplink.line[3].mode = tsplink.MODE_TRIGGER_OPEN_DRAIN
trigger.tsplinkout[3].logic = trigger.LOGIC_POSITIVE
Sets the trigger logic for
synchronization line 3 to output a
positive pulse.
Also see
trigger.tsplinkout[N].assert()
(on page 8-341)
(on page 8-348)
trigger.tsplinkout[N].pulsewidth
This attribute sets the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
TSP-Link line
N
reset
Configuration script
10e-6 (10
µ
s)
Usage
width
= trigger.tsplinkout[
N
].pulsewidth
trigger.tsplinkout[
N
].pulsewidth =
width
width
The pulse width (0.0 to 100 ks)
N