Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-228
DMM7510-901-01 Rev. B / May 2015
Example
:TRIGger:LOAD "LogicTrigger", 7, 2, 10, 0.001, "defbuffer1", ACT
Set up the template to use the external in line and wait for a pulse from the external in to trigger
measurements.
Pulse digital out line 2 when the measurement is complete. The external output line is also pulsed.
Make 10 measurements, with a delay of 1 ms before each measurement.
Store the measurements in
defbuffer1
.
Also see
None
:TRIGger:LOAD "LoopUntilEvent"
This command loads a predefined trigger model configuration that makes continuous measurements until the
specified event occurs.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <delay>
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <delay>,
"<readingBuffer>"
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <delay>,
"<readingBuffer>", <readingBlock>
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <clear>
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <clear>, <delay>
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <clear>, <delay>,
"<readingBuffer>"
:TRIGger:LOAD "LoopUntilEvent", <eventConstant>, <position>, <clear>, <delay>,
"<readingBuffer>", <readingBlock>
<eventConstant>
The event that ends infinite triggering or readings set to occur before the
trigger; see
Details
<position>
The number of readings to make in relation to the size of the reading
buffer; enter as a percentage
<clear>
To clear previously detected trigger events when entering the wait block
(default):
ENTer
To immediately act on any previously detected triggers and not clear them:
NEVer
<delay>
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
<readingBuffer>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is
specified,
defbuffer1
is used
<readingBlock>
Define a measure or digitize block for the trigger model; options are:
•
ACTive
: Add a measure or digitize block to the trigger model based
on the active function; if no option defined,
ACTive
is used
•
MEASure
: Adds a measure block to the trigger model
•
DIGitize
: Adds a digitize block to the trigger model