Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-212
DMM7510-901-01 Rev. B / May 2015
:TRIGger:EXTernal:OUT:STIMulus
This command selects the event that causes a trigger to be asserted on the external output line.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
NONE
Usage
:TRIGger:EXTernal:OUT:STIMulus <event>
:TRIGger:EXTernal:OUT:STIMulus?
<event>
The event to use as a stimulus; see
Details
Details
The trigger stimulus for the external output line can be set to one of the trigger events described in
the following table.
Trigger events
Event description
Event constant
No trigger event
NONE
Front-panel TRIGGER key press
DISPlay
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
A command interface trigger (bus trigger):
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
Trigger event blender
<n>
(1 or 2), which combines trigger events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Analog trigger
ATRigger
External in trigger
EXTernal
Example
:TRIG:EXT:OUT:STIMulus TIM3
Set the stimulus for the external output to be
the expiration of trigger timer 3.
Also see