Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-209
:TRIGger:EXTernal:IN:CLEar
This command clears the trigger event on the external in line.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:EXTernal:IN:CLEar
Details
The event detector of a trigger enters the detected state when an event is detected. This command
clears the event detector, discards the history, and clears the overrun status (sets the overrun status
to false).
Example
:TRIG:EXT:IN:CLE
Clears the trigger event detector on I/O line 2.
Also see
(on page 6-210)
:TRIGger:EXTernal:IN:EDGE
This command sets the type of edge that is detected as an input on the external in line.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
FALL
Usage
:TRIGger:EXTernal:IN:EDGE <detectedEdge>
:TRIGger:EXTernal:IN:EDGE?
<detectedEdge>
The trigger edge value:
•
Detect falling-edge triggers as inputs:
FALLing
•
Detect rising-edge triggers as inputs:
RISing
•
Detect either falling or rising-edge triggers as inputs:
EITHer
See
Details
for descriptions of values