Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-193
Example
TRIG:BLOC:CONF:PREV 14, "SETTINGS_LIST"
Set trigger block 14 to restore the settings
from the previous index that is stored in the
configuration list
SETTINGS_LIST
.
Also see
(on page 3-37)
:TRIGger:BLOCk:CONFig:RECall
This command recalls the system settings that are stored in a configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>"
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>", <index>
<blockNumber>
The sequence of the block in the trigger model
<configurationList>
A string that defines the configuration list to recall
<index>
The index in the configuration list to recall; defaults to 1 if not selected
Details
When the trigger model reaches a configuration recall building block, the settings in the specified
configuration list are recalled.
You can restore a specific set of configuration settings in the configuration list by defining the index.
Example
TRIG:BLOCK:CONF:RECALL 1, "SETTINGS_LIST", 1
Recall the settings in index 1 of the
SETTINGS_LIST
configuration list as block
1 of the trigger model.
Also see
(on page 3-37)