Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-129
STATus subsystem
The STATus subsystem controls the status registers of the instrument. For additional information on
the status model, see
(on page 1).
:STATus:CLEar
This function clears event registers and the event log.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:STATus:CLEar
Details
This command clears the event registers of the Questionable Event and Operation Event Register
set. It does not affect the Questionable Event Enable or Operation Event Enable registers.
Example
:STATus:CLEar
Clear the bits in the registers
Also see
(on page 2)
:STATus:OPERation:CONDition?
This command reads the Operation Event Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:OPERation:CONDition?
Details
This command reads the contents of the Operation Condition Register, which is one of the Operation
Event Registers.
For detail on interpreting the value of a register, see
(on page 14).
Example
:STAT:OPER:COND?
Returns the contents of the Operation
Condition Register.