Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-123
[:SENSe[1]]:DIGitize:FUNCtion[:ON]
This command selects which digitize function is active.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
NONE
Usage
[:SENSe[1]]:DIGitize:FUNCtion[:ON] "<function>"
[:SENSe[1]]:DIGitize:FUNCtion[:ON]?
<function>
A string that contains the measurement function to make active:
•
Current:
CURRent
•
Voltage:
VOLTage
Details
Set this command to the type of measurement you want to digitize.
Reading this command returns the digitize function that is presently active.
If you send the query when a measurement function is selected, the query returns
NONE
.
If a basic (non-digitize) measurement function is selected, this returns
NONE
. The none setting is
automatically made if you select a function with
[:SENSe[1]]:FUNCtion[:ON]
or through the
options from the front-panel Measure Functions tab.
Example
DIG:FUNC "VOLTage"
Make the digitize voltage function the active function.
Also see
(on page 6-124)