Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-92
DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:RELative
This command contains the relative offset value.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:RELative <n>
[:SENSe[1]]:<function>:RELative?
[:SENSe[1]]:<function>:RELative? DEFault
[:SENSe[1]]:<function>:RELative? MINimum
[:SENSe[1]]:<function>:RELative? MAXimum
<function>
The function to which the setting applies; see
Functions
<n>
The relative offset value; see
Details
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command specifies the relative offset value that can be applied to new measurements. When
relative offset is enabled, all subsequent measured readings are offset by the value that is set for this
command.
You can set this value, or have the instrument acquire a value. If the instrument acquires the value,
read this setting to return the value that was measured internally.
The ranges for the relative offset values for all functions are listed in the following table.