Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-84
DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:LINE:SYNC
This command determines if line synchronization is used during the measurement.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:LINE:SYNC <state>
[:SENSe[1]]:<function>:LINE:SYNC?
<function>
The function to which the setting applies; see
Functions
<state>
Disable:
OFF
or
0
Enable:
ON
or
1
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
When line synchronization is enabled, measurements are initiated at the first positive-going zero
crossing of the power line cycle after the trigger.
Example
CURR:LINE:SYNC ON
Turn on line synchronization when DC current is
measured.
Also see