Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-77
[:SENSe[1]]:<function>:COUPling:AC:FREQuency
This command allows you to optimize the amplitude to compensate for signal loss across the coupling capacitor
when AC coupling is selected.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1000
Usage
[:SENSe[1]]:<function>:AC:FREQuency <range>
[:SENSe[1]]:<function>:AC:FREQuency?
<function>
The function to which the setting applies; see
Functions
<range>
The frequency: 3 Hz to 1 MHz
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command is only used when the digitize coupling type is set to AC.
For example, if you are measuring a 50 Hz signal, you could set this to 50 Hz to compensate for
voltage drop across the coupling capacitor.
Example
DIG:VOLT:COUP AC
DIG:VOLT:COUP:AC:FILT FAST
DIG:VOLT:COUP:AC:FREQ 500
Set the coupling type to AC.
Set the filter to fast.
Set the frequency to 500 Hz.
Also see
[:SENSe[1]]:<function>:COUPling
(on page 6-75)