Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-61
[:SENSe[1]]:<function>:ATRigger:PULSe:LEVel
This command defines the pulse level that generates an analog trigger event.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:ATRigger:PULSe:LEVel <value>
[:SENSe[1]]:<function>:ATRigger:PULSe:LEVel?
<function>
The function to which the setting applies; see
Functions
<value>
The signal level
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
Only available when the analog trigger mode is set to pulse.
To use the analog trigger with the measure functions, a range must be set (you cannot use
autorange) and autozero must be disabled.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE PULSE
CURR:ATR:PULS:LEV 2.5
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to pulse.
Set the analog trigger level to 2.5 A.
Also see
[:SENSe[1]]:<function>:ATRigger:MODE
(on page 3-64)