Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-36
DMM7510-901-01 Rev. B / May 2015
:CALCulate[1]:<function>:MATH:PERCent
This command specifies the reference constant that is used when math operations are set to percent.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1
Usage
:CALCulate[1]:<function>:MATH:PERCent <n>
:CALCulate[1]:<function>:MATH:PERCent DEFault
:CALCulate[1]:<function>:MATH:PERCent MINimum
:CALCulate[1]:<function>:MATH:PERCent MAXimum
:CALCulate[1]:<function>:MATH:PERCent?
:CALCulate[1]:<function>:MATH:PERCent? DEFault
:CALCulate[1]:<function>:MATH:PERCent? MINimum
:CALCulate[1]:<function>:MATH:PERCent? MAXimum
<function>
The function to which the setting applies; see
Functions
<n>
The reference used when the math operation is set to percent; the range is -1e12 to
+1e12
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This is the constant that is used when the math operation is set to percent.
The percent math function displays measurements as percent deviation from a specified reference
constant. The percent calculation is:
Where:
•
Percent
is the result
•
Input
is the measurement (if relative offset is being used, this is the relative offset value)
•
Reference
is the user-specified constant
Example
CALC:VOLT:MATH:FORM PERC
CALC:VOLT:MATH:PERC 50
CALC:VOLT:MATH:STAT ON
Set the math operations for voltage to percent.
Set the percentage value to 50.
Enable math operations.
Also see
Calculations that you can apply to measurements
(on page 3-7)
:CALCulate[1]:<function>:MATH:FORMat
(on page 6-31)
:CALCulate[1]:<function>:MATH:STATe
(on page 6-37)