Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 3: Functions and features
DMM7510-901-01 Rev. B / May 2015
3-115
TSP-Link synchronization programming example
The programming example below illustrates how to set bit B1 of the TSP-Link digital I/O port high,
and then read the entire port value:
tsplink.line[1].mode = tsplink.MODE_DIGITAL_OPEN_DRAIN
-- Set bit B1 high.
tsplink.line[1].state = 1
-- Read I/O port.
data = tsplink.readport()
print(data)
The output would be similar to:
7
To read bit B1 only:
-- To read bit B1 only
data = tsplink.line[1].state
print(data)
The output would be similar to:
tsplink.STATE_HIGH
Using Model DMM7510 TSP-Link commands with other TSP-Link
products
If you are connecting the Model DMM7510 in a system with other TSP-Link products, be aware that
some of the TSP-Link commands may be different. You can use the earlier versions of the
commands, but be aware that they may not be supported in future versions of the product.
Commands that are the same in all TSP-Link products:
•
tsplink.group
•
tsplink.master
•
tsplink.node
•
tsplink.readport()
•
tsplink.state
•
tsplink.writeport()