2-17
Figure 2-5. Hardware Gate
Note:
Although DAS-8 Series boards do not provide a hardware-based
analog trigger, you can program an analog trigger through software, using
one of the analog input channels as the trigger channel. DriverLINX
provides functions for both an analog trigger and a digital trigger. Refer to
the DriverLINX
Appendix F: Configuration and Implementation Notes:
Keithley DAS-8
manual for more information.
Software starts
the operation
1st conversion
(external clock)
Internal Clock
Source
1st conversion
(internal clock)
2nd conversion
(internal clock)
3rd conversion
(internal clock)
4th conversion
(internal clock)
2nd conversion
(external clock)
External Clock
Source
3rd conversion
(external clock)
Gate Signal
Gate is high;
conversions occur
Gate is low;
conversions inhibited
. . . . . . . . . . . .
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com