Models 707B and 708B Switching Matrix Reference Manual
Section 7: Command reference
707B-901-01 Rev. A / August 2010
7-57
Also see
(on page 2-7)
(on page 7-55)
digio.trigger[N].assert()
This function asserts a trigger on one of the digital I/O lines.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
digio.trigger[N].assert()
N
Digital I/O trigger line (1 to 14)
Details
The set pulsewidth determines how long the trigger is asserted.
Example
digio.trigger[2].assert()
Asserts a trigger on digital I/O line 2
Also see
(on page 7-60)
digio.trigger[N].clear()
This function clears a trigger event on a digital I/O line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
digio.trigger[N].clear()
N
Digital I/O trigger line (1 to 14)
Details
The event detector of a trigger recalls if a trigger event has been detected since the last
digio.trigger[N].wait()
command. This function clears the event detector of a specified trigger line,
discards the previous history of the trigger line, and clears the
digio.trigger[N].overrun
attribute.