Section 7: Command reference
Models 707B and 708B Switching Matrix Reference Manual
7-50
707B-901-01 Rev. A / August 2010
Example 4
channel.setlabelrow("2B01", "Row2")
print(channel.getlabelrow("slot2"))
This example assumes a 7072 in slot 2.
Set the label to be Row2, which assigns the label
to all channels in the row. The output is now :
,,,,,,,,,,,,Row2,Row2,Row2,Row2,Row
2,Row2,Row2,Row2,Row2,Row2,Row2,
Row2,,,,,,,,,,,,,,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,
,,,,,,,,,,,,
Also see
(on page 7-32)
createconfigscript()
This function captures the present settings of the instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
createconfigscript(name)
name
The name of the script that will be created
Details
If name is set to
autoexec
, the autoexec script in the instrument will be replaced by the new configuration script.
If the name is set to the name of an existing script, the existing script will be overwritten.
Once created, the configuration script can be run and edited like any other script.
Example
createconfigscript("August2010")
Captures the present settings of the instrument
into a script named August2010
Also see
Save the present configuration